MINNEAPOLIS--(BUSINESS WIRE)--Feb. 11, 2005--Medtronic, Inc., (NYSE:MDT) today said it is voluntarily advising physicians about a potential battery shorting mechanism that may occur in a subset of implantable cardioverter-defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) models. In a letter to physicians, Medtronic reported that nine batteries (0.01 percent, or approximately 1 in 10,000) have experienced rapid battery depletion due to this shorting action. If shorting occurs, battery depletion can take place within a few hours to a few days, after which there is loss of device function. There are no reported patient injuries or deaths.
Devices with batteries manufactured between April 2001 and December 2003 may exhibit this shorting action. Potentially affected models are the Marquis(TM) VR/DR and Maximo(TM) VR/DR ICDs and the InSync I/II/III Marquis(TM) and InSync III Protect(TM) CRT-D devices. Approximately 75 percent of these devices were implanted in the United States. Patients with other models of Medtronic devices or these device models with batteries manufactured after December 2003 are not affected.
As part of the company's program to analyze products returned from physicians, Medtronic identified nine of 87,000 implanted devices (0.01 percent) with a battery design that exhibited this shorting mechanism. Based on highly accelerated company testing, Medtronic estimates that this rate may increase to between 0.2 percent and 1.5 percent over the second half of device life.
ICDs shock or pace the heart into normal rhythm after patients suffer rapid, life-threatening heart rhythm disturbances originating in the lower chambers of the heart that can lead to sudden cardiac arrest (SCA). CRT-Ds can also provide electrical pulses to the heart's two lower chambers to improve heart failure symptoms. An ICD or CRT-D device is surgically implanted in the chest in a procedure typically lasting one to two hours.
"We were able to identify this possible risk through our stringent product testing," said Steve Mahle, president of Medtronic Cardiac Rhythm Management. "Even though the potential for rapid battery depletion is extremely low, we see it as our obligation to alert all implanting physicians to the potential issue and provide ways to help them and their affected patients successfully manage the situation."
Medtronic began notifying physicians about the situation on Feb. 10, providing them with a list of potentially affected patients and recommending that physicians communicate with those patients and follow them closely. Medtronic is communicating this information so that physicians are aware of the potential issue and can manage it as they feel is appropriate with their individual patients.
Additional information about this potential issue is available for physicians and patients at 1-888-775-2702 (24/7) and www.medtronicinfo.com.
"I appreciate Medtronic's forthright communication about this issue, which will allow implanting physicians to manage this situation with each involved patient," said Dwight Reynolds, Chief of the Cardiovascular Section at the University of Oklahoma Health Sciences Center. "Physicians are in the best position to discuss the circumstances with device patients and ensure they have accurate information."
Medtronic, Inc., headquartered in Minneapolis, is the world's leading medical technology company, providing lifelong solutions for people with chronic disease. Its Internet address is www.medtronic.com.
Any forward-looking statements are subject to risks and uncertainties as described in Medtronic's Annual Report on Form 10-K for the year ended April 30, 2004. Actual results may differ materially from anticipated results.
--30--CR/ms*
CONTACT: Medtronic, Inc., Minneapolis Public Relations: Bob Hanvik, 763-505-2635 or Valerie Lind, 763-505-2631 or Investor Relations: Rachael Scherer, 763-505-2694 or Chris King, 763-505-2695
KEYWORD: MINNESOTA INDUSTRY KEYWORD: HARDWARE SOFTWARE MEDICAL DEVICES MEDICAL MANUFACTURING SOURCE: Medtronic, Inc.
Copyright Business Wire 2005
Wenn Sie mehr über das Thema Aktien erfahren wollen, finden Sie in unserem Ratgeber viele interessante Artikel dazu!
Jetzt informieren!
Nachrichten zu Medtronic PLCmehr Nachrichten
09.01.25 |
S&P 500-Wert Medtronic-Aktie: So viel Verlust hätte ein Investment in Medtronic von vor 3 Jahren bedeutet (finanzen.at) | |
08.01.25 |
Verluste in New York: Das macht der S&P 500 aktuell (finanzen.at) | |
08.01.25 |
Aufschläge in New York: Anleger lassen S&P 500 am Mittwochmittag steigen (finanzen.at) | |
08.01.25 |
Zurückhaltung in New York: S&P 500 präsentiert sich zum Start schwächer (finanzen.at) | |
02.01.25 |
S&P 500-Wert Medtronic-Aktie: So viel hätten Anleger an einem Medtronic-Investment von vor einem Jahr verloren (finanzen.at) | |
26.12.24 |
S&P 500-Papier Medtronic-Aktie: So viel Gewinn hätte eine Investition in Medtronic von vor 10 Jahren abgeworfen (finanzen.at) | |
19.12.24 |
S&P 500-Titel Medtronic-Aktie: So viel hätten Anleger an einem Medtronic-Investment von vor 5 Jahren verloren (finanzen.at) | |
12.12.24 |
S&P 500-Titel Medtronic-Aktie: So viel hätten Anleger mit einem Investment in Medtronic von vor 3 Jahren verloren (finanzen.at) |
Analysen zu Medtronic PLCmehr Analysen
Aktien in diesem Artikel
Medtronic PLC | 80,21 | -0,48% |
Indizes in diesem Artikel
S&P 500 | 5 918,25 | 0,16% | |
S&P 100 | 2 903,25 | 0,06% | |
NYSE US 100 | 16 447,86 | 0,76% |